Liquid chromatography parameters | |
---|---|
Chromatographic column | Waters BEH C18(50*2.1 mm, 1.8 μm) |
Mobile phase | Phase A: 0.1% formic acid aqueous solution Phase B: 0.1% formic acid acetonitrile solution |
Flow rate | 0.35 mL/min |
Column temperature | 40 ℃ |
Injection volume | 2 μL |
Elution gradient | 0.0 min A/B(90:10 V/V) 12.0 min A/B(10:90 V/V) 14.1 min A/B(90:10 V/V) 16.0 min A/B(90:10 V/V) |
Mass spectrometric parameters | |
Ion source | Electrospray ionization (ESI) |
Sheath gas | 40 arb |
Auxiliary gas | 10 arb |
Ion spray voltage | − 2800 V |
Temperature | 350 ℃ |
Ion transfer tube temperature | 320 ℃ |
Scan mode | Fullms-ms2 Mode |
Ion mode | Negative Ion Mode |
Data processing software | TraceFinder |